Noise sensitivity of 89 Zr-Immuno-PET radiomics based on count-reduced clinical images

EJNMMI Physics(2022)

引用 1|浏览22
暂无评分
摘要
Purpose Low photon count in 89 Zr-Immuno-PET results in images with a low signal-to-noise ratio (SNR). Since PET radiomics are sensitive to noise, this study focuses on the impact of noise on radiomic features from 89 Zr-Immuno-PET clinical images. We hypothesise that 89 Zr-Immuno-PET derived radiomic features have: (1) noise-induced variability affecting their precision and (2) noise-induced bias affecting their accuracy. This study aims to identify those features that are not or only minimally affected by noise in terms of precision and accuracy. Methods Count-split 89 Zr-Immuno-PET patient scans from previous studies with three different 89 Zr-labelled monoclonal antibodies were used to extract radiomic features at 50% (S50p) and 25% (S25p) of their original counts. Tumour lesions were manually delineated on the original full-count 89 Zr-Immuno-PET scans. Noise-induced variability and bias were assessed using intraclass correlation coefficient (ICC) and similarity distance metric (SDM), respectively. Based on the ICC and SDM values, the radiomic features were categorised as having poor [0, 0.5), moderate [0.5, 0.75), good [0.75, 0.9), or excellent [0.9, 1] precision and accuracy. The number of features classified into these categories was compared between the S50p and S25p images using Fisher’s exact test. All p values < 0.01 were considered statistically significant. Results For S50p, a total of 92% and 90% features were classified as having good or excellent ICC and SDM respectively, while for S25p, these decreased to 81% and 31%. In total, 148 features (31%) showed robustness to noise with good or moderate ICC and SDM in both S50p and S25p. The number of features classified into the four ICC and SDM categories between S50p and S25p was significantly different statistically. Conclusion Several radiomic features derived from low SNR 89 Zr-Immuno-PET images exhibit noise-induced variability and/or bias. However, 196 features (43%) that show minimal noise-induced variability and bias in S50p images have been identified. These features are less affected by noise and are, therefore, suitable candidates to be further studied as prognostic and predictive quantitative biomarkers in 89 Zr-Immuno-PET studies.
更多
查看译文
关键词
89Zr-Immuno PET,Radiomics,Noise,Precision,Bias,Repeatability,Reproducibility
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要