Reliability enhancement of non-contact fever screening technology (thermography, etc.) for quarantine inspection

Measurement: Sensors(2021)

引用 2|浏览0
暂无评分
摘要
We have been developing the technology for reliable non-contact body temperature measurement, which contributes to the prevention of new infectious diseases. As a part of this study, we are developing a highly accurate on-site temperature reference with a planar blackbody (reference infrared radiator) that can correct fever-screening thermography. We have established a process to manufacture a planar blackbody material offering a high emissivity of >0.99, which would be less affected by surroundings (background infrared radiation, etc.) when correcting and evaluating thermographic cameras. The target temperature setting accuracy of the planar blackbody radiator would be < 0.2 °C. Thereby the error factors of fever-screening thermography including device performance (stability, response uniformity, etc.), target parameters (emissivity, distance, etc.), and ambient environment (temperature, etc.) would be inspected and corrected in more detail.
更多
查看译文
关键词
Fever screening,Non-contact,Thermography,Calibration,Planar blackbody