Non-contact automatic vital signs monitoring of neonates in NICU using video camera imaging

COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING-IMAGING AND VISUALIZATION(2023)

引用 2|浏览1
暂无评分
摘要
Non-contact vital signs monitoring is desirable in both clinical and non-clinical sectors. However, there are limited camera-based studies to monitor the vital signs of neonatal infants in a clinical setting. In this study, a digital camera was used to remotely monitor the heart rate (HR) and respiratory rate (RR) of seven infants at Flinders Medical Centre Neonatal Intensive Care Unit (NICU). The region-of-interest (ROI) was automatically selected based on a skin detection method using skin hue. The experimental results have been validated with the reference data obtained from an ECG monitor, and they showed a strong correlation using the Pearson correlation coefficient, (PCC) .9836 and .9426 for HR and RR, respectively, and a lower error rate with RMSE 2.66 beats/min and 2.93 breaths/min between measured data and reference data. A Bland Altman analysis of the data also presented a close correlation between measured data and reference data for both HR and RR with a mean bias .69 beats/min and .28 breaths/min, and the lower and upper limit of agreement -4.4 to +5.8 beats/min and -5.5 to +6.1 breaths/min, respectively. Therefore, video camera imaging may replace conventional monitoring in NICU and has potential applications in other contexts such as home health monitoring.
更多
查看译文
关键词
Heart rate, respiratory rate, neonates, NICU, digital camera, skin detection