In our previous studies, we have demonstrated the feasibility of characterizing intestinal inflammation and fibrosis using endoscopic photoacoustic imaging. Purposed at te clinical translation of the imaging technology, we developed a photoacoustic/ultrasound imaging probe by integrating a miniaturized ultrasound array and an angle-tipped optical fiber in a hydrostatic balloon catheter. When collapsed, the catheter probe may potentially be compatible with a clinical ileo-colonoscope. In addition, the flexible surface of the hydrostatic balloon allows for acoustic coupling at the uneven surfaces of the gas-filled intestine. Tissue phantom studies show that the catheter probe possesses an imaging penetration of at least 12 mm. Experiments with a rabbit model in vivo validated the probe in differentiating normal, acute and chronic conditions in intestinal obstruction.