Personal dosimetry for positron emitters, and occupational exposures from clinical use of gallium-68

JOURNAL OF RADIOLOGICAL PROTECTION(2022)

引用 2|浏览7
暂无评分
摘要
The current status and issues regarding positron dosimetry in nuclear medicine are summarized. The suitability of the United Kingdom Health Security Agency extremity and eye beta-gamma personal thermoluminescence dosemeters are then considered. Monte Carlo modelling is performed to determine their responses and derive sets of calibration factors, along with H (p)(0.07) and H (p)(3) conversion coefficients, for carbon-11, nitrogen-13, oxygen-15, fluorine-18 and gallium-68 sources, which are commonly used in positron emission tomography (PET) computed tomography; data for these isotopes is assumed extrapolatable to other positron sources. It is found that the dosemeters are adequate for assessing exposures to PET radionuclides, even if their routine calibrations to caesium-137 were maintained. An idealized set of measurements representing gallium-68 exposure scenarios is then described, including reproducible mock-ups of individuals manipulating vials and syringes. Finally, a short case-study is presented that explores occupational doses during routine clinical use of gallium-68. The extremity dosemeter results demonstrated significant variations dependent upon the exposure conditions, with some seen to be comparatively large; whole-body and eye dose rates per activity were found to be lower. The importance of routine dose monitoring of workers is emphasized, with the need for a longer-termed follow-up study demonstrated.
更多
查看译文
关键词
personal dosimetry, positron, PET-CT, calibration, Monte Carlo, occupational exposures, medical dosimetry
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要