Reconstruction of Mandibular Contour Defect Using Patient-Specific Titanium Implant Manufactured by Selective Laser Melting Method

JOURNAL OF CRANIOFACIAL SURGERY(2022)

引用 3|浏览0
暂无评分
摘要
The purpose of this study was to evaluate the surgical accuracy and postoperative stability of patient-specific titanium implants (PSTIs) manufactured by the selective laser melting method and applied for mandibular contour reconstruction. For 2 patients who showed asymmetry of the mandibular angle after mandibuloplasty, including angle reduction, reconstructive surgeries of the mandibular contour defects were performed using PSTI. Patient-specific titanium implant was three-dimensional designed using a mirror image similar to the shape of the contralateral side, and 3 screw holes were formed, avoiding the inferior alveolar nerve. Patient-specific titanium implant was applied intraorally, and screw fixation was performed via a transbuccal approach. Surgical accuracy and postoperative stability were evaluated by comparing preoperative three-dimensional design with immediate postoperative computed tomography (CT), and immediate postoperative CT with postoperative 6-month CT, respectively. Both patients were satisfied with the surgical results, and no complications were observed. Surgical accuracy was defined as a mean Hausdorff distance <0.4mm, and postoperative stability was defined as a mean Hausdorff distance <0.3mm. Our results suggest that PSTI by the selective laser melting method for augmentation of the mandibular border is useful as an additional cosmetic surgery.
更多
查看译文
关键词
3D printing, accuracy, mandible, patient-specific titanium implant, reconstruction