Early diagnosis of cardiovascular diseases (CVDs) has the potential to save millions of lives each year. Sensitive quantitative measurement of blood cardiac troponin I (cTnI) is required for early diagnosis of CVDs. Porous graphitic carbon nitride (PCN) nanomaterials integrated biosensor that detects picogram/mL concentrations of cTnI is reported. PCN is an updated version of graphitic carbon nitride (GCN) with improved porosity and electronic structure. A rapid two-step chemical method was described to synthesize PCN and gold nanoparticles functionalized PCN (PCN-AuNPs). The modification of the electrode surface with PCN materials had a significant impact on the electrochemically active surface area (EASA), interfacial electron transport, aptamer immobilization, and biosensing performance. The use of PCN nanomaterial in cTnI aptasensing resulted in a 3.2-fold increase in signal amplification. PCN-AuNPs found practically applicable in human blood serum (cTnI-spiked). Furthermore, a low-cost and user-friendly sensing platform has been demonstrated by integrating a PCN-AuNPs aptasensor, a custom-made miniaturized potentiostat, and a smartphone, which provided rapid, sensitive point-of-care (PoC) analysis of cTnI. Highly sensitive detection limit (0.01 pg/mL), rapid analysis time (2 min), and small sample volume (20 μL) are the other advantages of this nano-biosensor.