Study on the Blood Flow Characteristics of Venous Needle Retention with Different Super-Hydrophobic Surface Structures.

Zhun Yu, Lei Liu,Yongzhi Deng, Xiaowen Zhang,Chao Yu

Medical & biological engineering & computing(2023)

引用 1|浏览19
暂无评分
摘要
A venous retention needle, as an implanted device, is very likely to cause thrombosis. In view of the thrombosis phenomenon caused by retention needles, this paper compares the influence of different superhydrophobic surface retentions on blood flow. Compared with other superhydrophobic bulges, the fluid velocity of the four-prism bulge is the highest (0.08 m/s), and the vorticity and shear force of the hemispherical bulge are higher. A large number of vortices can inhibit thrombosis better. The tire vortices generated in the superhydrophobic convex grooves are important vortices to inhibit thrombosis. The enhancement and development of the tire vortex weakens the resistance near the wall of the needle and reduces the probability of platelet aggregation. The superhydrophobic surface structure studied in this paper can not only provide guidance for the design of venous retention needles with better performance but also provide corresponding technical support for the development of human implantation devices.
更多
查看译文
关键词
Vein retention needle,Blood,Super-hydrophobic surface,Computational fluid dynamics