Novel screw fixation placement configuration for the treatment of Pauwels type III femoral neck fractures: a finite element analysis

Medical & Biological Engineering & Computing(2023)

引用 1|浏览1
暂无评分
摘要
Verticality of transcervical hip fractures in young patients is usually connected with typically high-energy fractures which are known as Pauwels type III. Artificial femoral head replacement surgery is mostly not considered for treating femoral neck fractures in such patients. The commonly used devices for the fixation of vertical femoral neck fractures are multiple screws or a sliding hip screw with or without an antirotation screw. Size, location and length of the screws are the most effective parameters in terms of the structural performance of internal fixation implants, but the optimal configuration of the screws is necessary to be investigated to direct the clinical practice. The aim of this study is to compare the biomechanical stability of the standard inverted triangle configuration with the various newly proposed x -crossed screw configurations. FEA simulations carried out in this study demonstrated that using an x -crossed-right assembly in treating Pauwels type III femoral neck fractures satisfies the biomechanical stability in terms of maximum von Mises stresses and maximum femoral head displacement. However, in terms of maximum relative neck fracture displacement, the x -crossed-right assembly would not entirely suffice the desired biomechanical stability. Therefore, using an x -crossed screw assembly in treating femoral neck fractures would provide the needed biomechanical stability. Graphical Abstract
更多
查看译文
关键词
Biomechanical stability,Femoral neck fracture,Finite element analysis,Inverted triangle,Pauwels type III
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要