Validation of combined carcinoembryonic antigen and glucose testing in pancreatic cyst fluid to differentiate mucinous from non-mucinous cysts

Surgical endoscopy(2023)

引用 0|浏览16
暂无评分
摘要
Background More accurate diagnosis of mucinous cysts will reduce the risk of unnecessary pancreatic surgery. Carcinoembryonic antigen (CEA) and glucose in pancreatic cyst fluid (PCF) can differentiate mucinous from non-mucinous pancreatic cystic neoplasms (PCN). The current study assessed the value of combined CEA and glucose testing in PCF. Methods Cross-sectional validation study including prospectively collected PCF from patients undergoing endoscopic ultrasonography-guided fine-needle aspiration (EUS-FNA) and pancreatic surgery. We performed laboratory measurements for CEA and glucose and measured glucose levels by a hand glucometer. Primary outcome was diagnostic accuracy evaluated by receiver operator curves (ROC), sensitivity, specificity, positive, and negative predictive value (PPV, NPV). Results Overall, PCF was collected from 63 patients, including 33 (52%) with mucinous and 30 (48%) with non-mucinous PCN. Histopathology ( n = 36; 57%), cytopathology ( n = 2; 3%), or clinical and/or radiological diagnosis ( n = 25; 40%) was used as reference standard. Combined CEA (cut-off ≥ 192 ng/ml) and laboratory glucose testing (cut-off ≤ 50 mg/dL) reached 92% specificity and 48% sensitivity, whereas either positive CEA (cut-off ≥ 20 ng/ml) or glucose testing (cut-off ≤ 50 mg/dL) showed 97% sensitivity and 50% specificity. Sensitivity and specificity were 80% and 68% for CEA ≥ 20 ng/mL versus 50% and 93% for CEA ≥ 192 ng/mL (the conventional cut-off level). Laboratory and glucometer glucose both reached 100% sensitivity and 60% and 45% specificity, respectively. None of the biomarkers and cut-offs reached a PPV exceeding 90%, whereas both glucose measurements had a NPV of 100% (i.e., high glucose excludes a mucinous cyst). Conclusion Combined CEA and glucose testing in PCF reached high specificity and sensitivity for differentiating mucinous from non-mucinous PCN. Glucose testing, whether alone or combined with the new CEA cut-off (≥ 20 ng/mL), reached > 95% sensitivity for mucinous cysts, whereas only glucose reached a NPV > 95%. Graphical abstract
更多
查看译文
关键词
Carcinoembryonic antigen (CEA),Diagnostic accuracy,Endoscopy: Endoscopic ultrasound diagnostic and therapeutic,Glucose,Hepatobiliary tract and spleen,Pancreatic cyst fluid,Pancreatic cystic neoplasms
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要