Developing a rapid antibody-based detection method is of great importance for preventing and controlling the spread of coronavirus disease 2019 (COVID-19). Among the antibody-based methods for point-of-care (POC) detection, lateral flow immunoassay (LFIA) is the most widely used. However, LFIA still has the disadvantage of low sensitivity. In this work, an ReSe2 nanosheet with a thickness of 10-20 nm was prepared by liquid exfoliation and applied as the label in a photothermal LFIA due to its high photothermal conversion efficiency and high photothermal stability. An integrated detection device was introduced for rapid, on-site, and highly sensitive assay of the human antisevere acute respiratory syndrome coronavirus 2 (SARS-CoV-2) Spike (S) protein IgG antibodies. The device mainly included a rhenium diselenide (ReSe2) nanosheet-based photothermal LFIA, a portable laser, and a smartphone with a portable thermal imager, which was used to record and analyze the thermal signal of the LFIA test zone. The human anti-SARS-COV-2 S protein IgG antibodies in buffer solution can be detected in a portable box within 10 min, with a thermal signal detection limit of 0.86 ng mL-1, which was 108-fold lower than that of the colorimetric signal. The integrated device can detect values as low as 2.76 ng mL-1 of the human anti-SARS-COV-2 S protein IgG antibodies in 50% serum. The integrated device showed great potential for rapid and home self-testing diagnosis of COVID-19.