Elevated methemoglobin levels in patients treated with high-dose hydroxocobalamin

Martinus Dyrud,Jianli Niu, Lisa Kohler

LABORATORY MEDICINE(2024)

引用 0|浏览0
暂无评分
摘要
Objective The aim of this study was to assess the impact of hydroxocobalamin (OHCbl) infusion on arterial blood gas and oximetry values in patients with vasoplegic syndrome. Methods Blood samples collected from 95 patients receiving OHCbl infusion were assayed using the ABL90 FLEX Plus blood gas analyzer for the concentration of methemoglobin (MetHb), total hemoglobin (tHb), carboxyhemoglobin (COHb), arterial oxygen saturation (SaO(2)), arterial oxygen partial pressure (PaO2), and arterial carbon dioxide partial pressure (PaCO2). Interference of OHCbl on these variables was evaluated using the measured difference between the preinfusion and postinfusion samples. Results Blood MetHb (%) measured after the infusion of OHCbl (5g) were significantly higher than the baseline levels, with a median of 4.8 (IQR, 3.0-6.5) versus 1.0 (IQR, 1.0-1.2) (P < .001). Blood COHb (%) increased from a median of 1.3 (IQR, 1.0-1.8) to 1.7 (IQR, 1.3-2.2) (P < .001) following the OHCbl infusion. No differences were seen in median levels of tHb, PaO2, PaCO2, and SaO(2) between pre- and post-OHCbl treatment. Conclusion The presence of OHCbl in blood clearly interfered with the oximetry measurements of the hemoglobin component fractions by falsely increasing the levels of MetHb and COHb. Blood levels of MetHb and COHb cannot be reliably determined by the co-oximetry when OHCbl is known or suspected.
更多
查看译文
关键词
hydroxocobalamin,methemoglobin,carboxyhemoglobin,arterial blood gas analysis,oximetry assay