Technique for removal of a ganciclovir implant dislocated in the vitreous cavity

RETINAL Cases & Brief Reports(2022)

引用 0|浏览2
暂无评分
摘要
Purpose: To report a technique to remove a dislocated ganciclovir implant in the vitreous cavity. Methods: Retrospective case series. Patients: Two patients with dislocated ganciclovir implants in the vitreous cavity. Results: A 6mm pars plana incision was made, the soft tip was used to elevate the implant behind the intraocular lens and then 0.12mm forceps were used to externalize the implant. The implant was successfully removed in both patients. Discussion: Removal of a dislocated ganciclovir implant with its encasing strut can be effectively retrieved using a bimanual approach.
更多
查看译文
关键词
ganciclovir implant,removal
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要