Evaluation of the venting principle to reduce start-up delays in syringe infusion pumps used for microinfusions.

Journal of clinical monitoring and computing(2023)

引用 0|浏览18
暂无评分
摘要
Start-up delays of syringe pump assemblies can impede the timely commencement of an effective drug therapy when using microinfusions in hemodynamically unstable patients. The application of the venting principle has been proposed to eliminate start-up delays in syringe pump assemblies. However, effectively delivered infusion volumes using this strategy have so far not been measured. This invitro study used two experimental setups to measure the effect of the venting principle compared to a standard non-venting approach on delivered start-up infusion volumes at various timepoints, backflow volumes, flow inversion and zero drug delivery times by means of liquid flow measurements at flow rates of 0.5, 1.0 and 2.0 mL/h. Measured delivered initial start-up volumes were negative with all flow rates in the vented and non-vented setup. Maximum backflow volumes were 1.8 [95% CI 1.6 to 2.3] times larger in the vented setup compared to the non-vented setup (p < 0.0001). Conversely, times until flow inversion were 1.5 [95% CI 1.1 to 2.9] times shorter in the vented setup (p < 0.002). This led to comparable zero drug delivery times between the two setups (p = 0.294). Start-up times as defined by the achievement of at least 90% of steady state flow rate were achieved faster with the vented setup (p < 0.0001), but this was counteracted by the increased backflow volumes. The application of the venting principle to the start-up of microinfusions does not improve the timely delivery of drugs to the patient since the faster start-up times are counteracted by higher backflow volumes when opening the three-way stopcock.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要