Reproducible spectral CT thermometry with liver-mimicking phantoms for image-guided thermal ablation

PHYSICS IN MEDICINE AND BIOLOGY(2024)

引用 0|浏览11
暂无评分
摘要
Objectives. Evaluate the reproducibility, temperature tolerance, and radiation dose requirements of spectral CT thermometry in tissue-mimicking phantoms to establish its utility for non-invasive temperature monitoring of thermal ablations. Methods. Three liver mimicking phantoms embedded with temperature sensors were individually scanned with a dual-layer spectral CT at different radiation dose levels during heating (35 degrees C-80 degrees C). Physical density maps were reconstructed from spectral results using varying reconstruction parameters. Thermal volumetric expansion was then measured at each temperature sensor every 5 degrees C in order to establish a correlation between physical density and temperature. Linear regressions were applied based on thermal volumetric expansion for each phantom, and coefficient of variation for fit parameters was calculated to characterize reproducibility of spectral CT thermometry. Additionally, temperature tolerance was determined to evaluate effects of acquisition and reconstruction parameters. The resulting minimum radiation dose to meet the clinical temperature accuracy requirement was determined for each slice thickness with and without additional denoising. Results. Thermal volumetric expansion was robustly replicated in all three phantoms, with a correlation coefficient variation of only 0.43%. Similarly, the coefficient of variation for the slope and intercept were 9.6% and 0.08%, respectively, indicating reproducibility of the spectral CT thermometry. Temperature tolerance ranged from 2 degrees C to 23 degrees C, decreasing with increased radiation dose, slice thickness, and iterative reconstruction level. To meet the clinical requirement for temperature tolerance, the minimum required radiation dose ranged from 20, 30, and 57 mGy for slice thickness of 2, 3, and 5 mm, respectively, but was reduced to 2 mGy with additional denoising. Conclusions. Spectral CT thermometry demonstrated reproducibility across three liver-mimicking phantoms and illustrated the clinical requirement for temperature tolerance can be met for different slice thicknesses. The reproducibility and temperature accuracy of spectral CT thermometry enable its clinical application for non-invasive temperature monitoring of thermal ablation.
更多
查看译文
关键词
spectral computed tomography,computed tomography,thermometry,CT thermometry,thermal ablation,interventional radiology,microwave ablation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要