In vitro investigation of stone ablation efficiency, char formation, spark generation, and damage mechanism produced by thulium fiber laser

Urolithiasis(2023)

引用 1|浏览7
暂无评分
摘要
To investigate stone ablation characteristics of thulium fiber laser (TFL), BegoStone phantoms were spot-treated in water at various fiber tip-to-stone standoff distances (SDs, 0.5 ~ 2 mm) over a broad range of pulse energy (E p , 0.2 ~ 2 J), frequency (F, 5 ~ 150 Hz), and power (P, 10 ~ 30 W) settings. In general, the ablation speed (mm 3 /s) in BegoStone decreased with SD and increased with E p , reaching a peak around 0.8 ~ 1.0 J. Additional experiments with calcium phosphate (CaP), uric acid (UA), and calcium oxalate monohydrate (COM) stones were conducted under two distinctly different settings: 0.2 J/100 Hz and 0.8 J/12 Hz. The concomitant bubble dynamics, spark generation and pressure transients were analyzed. Higher ablation speeds were consistently produced at 0.8 J/12 Hz than at 0.2 J/100 Hz, with CaP stones most difficult yet COM and UA stones easier to ablate. Charring was mostly observed in CaP stones at 0.2 J/100 Hz, accompanied by strong spark-generation, explosive combustion, and diminished pressure transients, but not at 0.8 J/12 Hz. By treating stones in parallel fiber orientation and leveraging the proximity effect of a ureteroscope, the contribution of bubble collapse to stone ablation was found to be substantial (16% ~ 59%) at 0.8 J/12 Hz, but not at 0.2 J/100 Hz. Overall, TFL ablation efficiency is significantly better at high E p /low F setting, attributable to increased cavitation damage with less char formation.
更多
查看译文
关键词
Thulium Fiber laser,Stone ablation,Charring,Spark Generation,Cavitation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要