Development of an event-by-event based Li-ZnS(Ag) neutron imaging detector with selective neutron detection capability

APPLIED RADIATION AND ISOTOPES(2024)

引用 0|浏览1
暂无评分
摘要
High sensitivity and high resolution is desired in such technologies as neutron radiography. However, the contamination of gamma photons in neutron images decreases the accuracy of neutron radiography. To solve this problem, we developed an event-by-event based neutron imaging system that can selectively detect neutrons. The developed neutron imaging system consists of an Li-ZnS(Ag) scintillator plate optically coupled to a flat panel photomultiplier tube (FP-PMT) with a light guide. Scintillation light emitted from the Li-ZnS(Ag) by the interaction with neutron-induced particles is used to calculate the position based on the center of mass calculations. The spatial resolution of the neutron imaging detector is similar to 2.3 mm FWHM, and the sensitivity for Cf-252 at 2 cm from the source with 2-cm-thick polystyrene is 20 cps/MBq. Background count fraction from Co-60 gamma photons was 0.2 %. For various types of neutron absorption phantoms, high-contrast and high-resolution neutron images are obtained with the developed neutron imaging detector using a(252)Cf neutron source.
更多
查看译文
关键词
Neutron,Induced triton and alpha particle,Li-ZnS(Ag),FP-PMT