Label-free Microfluidic Chip for Segregation and Recovery of Circulating Leukemia Cells: Clinical Applications in Acute Myeloid Leukemia

Dongfang Ouyang, Ningxin Ye, Yue Jiang, Yiyang Wang,Lina Hu, Shuen Chao,Martin Yarmush, Memet Tuner,Yonghua Li,Bin Tang

BIOMEDICAL MICRODEVICES(2024)

引用 0|浏览6
暂无评分
摘要
We present a label-free microfluidic chip for the segregation of circulating leukemia cells (CLCs) from blood samples, with a focus on its clinical applications in Acute Myeloid Leukemia (AML). The microfluidic chip achieved an approximate capture efficiency of 92%. The study analyzed a comprehensive set of 66 blood specimens from AML patients in different disease stages, including newly diagnosed and relapsing cases, patients in complete remission, and those in partial remission. The results showed a significant difference in CLC counts between active disease stages and remission stages (p < 0.0001), with a proposed threshold of 5 CLCs to differentiate between the two. The microfluidic chip exhibited a sensitivity of 95.4% and specificity of 100% in predicting disease recurrence. Additionally, the captured CLCs were subjected to downstream molecular analysis using droplet digital PCR, allowing for the identification of genetic mutations associated with AML. Comparative analysis with bone marrow aspirate processing by FACS demonstrated the reliability and accuracy of the microfluidic chip in tracking disease burden, with highly agreement results obtained between the two methods. The non-invasive nature of the microfluidic chip and its ability to provide real-time insights into disease progression make it a promising tool for the proactive monitoring and personalized patient care of AML.
更多
查看译文
关键词
Microfluidic chip,Acute myeloid leukemia,Liquid biopsy,Circulating biomarkers,Diagnosis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要