Effect of decentration on the quality of vision in two aspheric posterior chamber intraocular lenses: A contralateral eye study.

Anchal Thakur, Sushant Adiga,Chintan Malhotra,Kulbhushan Sachdeva, Shubhi Singh,Arun Kumar Jain

Indian journal of ophthalmology(2024)

引用 0|浏览1
暂无评分
摘要
PURPOSE:To perform an intraindividual comparison of the quality of vision and the effect of decentration between two aspheric intraocular lenses: aspheric balanced curve (ABC) design Vivinex iSert XY1 (Hoya Surgical Optics, Singapore) and anterior aspheric design Tecnis ZCB00 (Abbott Medical Optics, CA). SETTING:Tertiary Eye Care Centre. DESIGN:Prospective, randomized comparative study using a random number table. METHODS:Thirty patients were randomized to the implantation of Vivinex iSert XY1 in one eye and Tecnis ZCB00 in the contralateral eye. Then, 12 weeks postoperatively, a laser ray-tracing aberrometer was used to evaluate the visual Strehl ratio, higher-order aberrations (HOA), decentration of IOL from the visual axis and geometric axis, angle alpha, and angle kappa. Contrast sensitivity was measured using the functional visual analyzer. RESULTS:The visual internal Strehl ratio was higher ( P < 0.05) at all pupil sizes and the spherical aberrations values were lower ( P < 0.05) at larger pupil sizes (5 mm and 6 mm) in the Vivinex group. The mean decentration from the visual axis in the Vivinex group was significantly more than that in the Tecnis group ( P < 0.01). With an increase in angle alpha, there was a greater decline in the visual Strehl ratio for Tecnis; however, there was a statistically significant decline at 3 mm pupil size for Tecnis ZCB00 ( P = 0.028). The contrast sensitivity was similar for both IOLs. CONCLUSION:In comparison to an anterior aspheric design IOL, the ABC design IOL yielded better quality of vision, neutralized spherical aberrations to a larger extent, and provided a relatively superior quality of vision with decentration.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要