The Detection of Trace Metal Contaminants in Organic Products Using Ion Current Rectifying Quartz Nanopipettes

ANALYTICAL CHEMISTRY(2024)

引用 0|浏览1
暂无评分
摘要
While ion current rectification (ICR) in aprotic solvent has been fundamentally studied, its application in sensing devices lacks exploration. The development of sensors operable in these solvents is highly beneficial to the chemical industry, where polar aprotic solvents, such as acetonitrile, are widely used. Currently, this industry relies on the use of inductively coupled plasma mass spectrometry (ICP-MS) and optical emission spectroscopy (OES) for the detection of metal contamination in organic products. Herein, we present the detection of trace amounts of Pd2+ and Co2+ using ion current rectification, in cyclam-functionalized quartz nanopipettes, with tetraethylammonium tetrafluoroborate (TEATFB) in MeCN as supporting electrolyte. This methodology is employed to determine the concentration of Pd in organic products, before and after purification by Celite filtration and column chromatography, obtaining comparable results to ICP-MS within minutes and without complex sample preparation. Finite element simulations are used to support our experimental findings, which reveal that the formation of double-junction diodes in the nanopore enables trace detection of these metals, with a significant response from baseline even at picomolar concentrations.
更多