A comparison of respiratory oscillometry and spirometry in idiopathic pulmonary fibrosis: performance time, symptom burden and test-retest reliability

ERJ Open Research(2024)

引用 0|浏览4
暂无评分
摘要
Study questionIn large multinational patient surveys, spirometry (which requires repeated, reproducible maximal efforts) can be associated with cough, breathlessness and tiredness, particularly in those with idiopathic pulmonary fibrosis (IPF). Oscillometry is an effort-independent test of airways resistance and reactance. We hypothesised that oscillometry would take less time to perform and would be associated with reduced symptom burden than spirometry.Patients and methodsSpirometry and oscillometry were performed in sixty-six participants with IPF and repeated two weeks later. We compared time taken to perform tests, symptom burden and test-retest reliability with Bland-Altman plots and intraclass correlation coefficients (ICCs).ResultsOscillometry took significantly less time to perform than spirometry (mean (99% CI) −4.5 (−6.0 to −3.0) minutes) and was associated with lower symptom burden scores for cough (−1.3 (−1.7 to −0.8)), breathlessness (−1.0 (−1.4 to −0.5)), and tiredness (−0.5 (−0.9 to −0.2)). On Bland-Altman analysis, all measures showed good agreement, with narrow limits of agreement and the mean bias lying close to 0 in all cases. The ICC for Forced expiratory volume in one second (FEV1) and Forced Vital Capacity (FVC) were 0.94 and 0.89, and ranged between 0.70 and 0.90 for oscillometry measures.ConclusionOscillometry is quicker to perform and provokes less symptoms than spirometry in patients with IPF.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要