Digital breast tomosynthesis (DBT) is an imaging modality in which tomographic sections of the breast are generated from a limited range of x-ray tube angles One drawback of DBT is resolution loss in the oblique projection images The purpose of this work is to extend Swank's formulation of the transfer functions of turbid granular phosphors to oblique x-ray incidence, using the diffusion approximation to the Boltzmann equation to model the spread of light in the phosphor As expected, the modulation transfer function (MTF) and noise power spectra (NPS) are found to decrease with projection angle regardless of frequency By contrast, the dependence of detective quantum efficiency (DQE) on projection angle is frequency dependent DQE increases with projection angle at low frequencies, and only decreases with projection angle at high frequencies Importantly, the x-ray quantum detection efficiency (AQ) and the Swank information factor (AS) are also found to be angularly dependent.
更多
查看译文
关键词
Digital breast tomosynthesis (DBT),oblique x-ray incidence,turbid granular phosphor,optical transfer function (OTF),modulation transfer function (MTF),noise power spectra (NPS),detective quantum efficiency (DQE),Swank information factor (A S)