Computed tomography attenuation correction (CTAC) is commonly used in cardiac SPECT imaging to reduce soft-tissue attenuation artifacts. However, CTAC is prone to inaccuracies due to CT artifacts and SPECT-CT mismatch, along with additional radiation exposure to patients. Thus, these limitations have led to increasing interest in CT-free AC, with deep learning (DL) offering promising solutions. We proposed a new DL-based CT-free AC methods for cardiac SPECT. We developed a feature alignment attenuation correction network (FA-ACNet) based on the 3D U-Net framework to generate predicted DL-based AC SPECT (Deep AC). The network was trained on 167 cardiac SPECT/CT studies using 5-fold cross validation and tested in an independent testing set (n = 35), with CTAC serving as the reference. During training, multi-scale features from non-attenuation-corrected (NAC) SPECT and CT were processed separately and then aligned with the encoded features from NAC SPECT using adversarial learning and distance metric learning techniques. The performance of FA-ACNet was evaluated using mean square error (MSE), structural similarity index (SSIM) and peak signal-to-noise ratio (PSNR). Additionally, semi-quantitative evaluation of Deep AC images was performed and compared to CTAC using Bland-Altman plots. FA-ACNet achieved an MSE of 16.94 ± 2.03 × 10− 6, SSIM of 0.9955 ± 0.0006 and PSNR of 43.73 ± 0.50 after 5-fold cross validation. Compared to U-Net, MSE and PSNR improved by aligning multi-scale features from NAC SPECT and CT with those from NAC SPECT. In the testing set, FA-ACNet achieved an MSE of 11.98 × 10− 6, SSIM of 0.9976 and PSNR of 45.54. The 95
更多
查看译文
关键词
Deep learning,SPECT,Myocardial perfusion imaging,Attenuation correction