Towards a Phantom-Based Quantitative Assessment of Dose Reduction and Image-Quality Trade-Offs in Half- and Full-Rotation Dental CBCT Protocols: a Thermoluminescent Dosimetry Approach with Clinical Insight
This study aimed to compare full- (360°) and half-rotation (180°) dental cone-beam computed tomography (CBCT) protocols in terms of effective dose (ED) and quantitative image quality for the maxilla and mandible regions. An Alderson Radiation Therapy phantom was imaged using a Hyperion X9 Pro (Cefla, Imola) CBCT device with a 10 × 6 cm field of view. Both anatomical regions were acquired using full- and half-rotation protocols. A total of 67 thermoluminescence dosimeters were positioned in the phantom for ED calculations. Quantitative image-quality assessment was performed using axial slices extracted from each volume, and signal-to-noise ratio (SNRs) and contrast-to-noise ratio (CNRs) were calculated. The EDs were 406.33 µSv (full-rotation) and 208.29 µSv (half-rotation) for the maxilla, and 248.94 µSv (full-rotation) and 73.63 µSv (half-rotation) for the mandible. These doses corresponded to dose reductions of 48.74