Simulation of Estimating the Blood Pressure Using an Arterial Pressure-Volume Model

ICCIT '07 Proceedings of the 2007 International Conference on Convergence Information Technology(2007)

引用 5|浏览1
暂无评分
摘要
The abstract is to be in an analysis on the conventional the blood pressure estimation method using an oscillometric sphygmomanometer was performed through a computer simulation using an arterial pressure-volume (APV) model. Traditionally, the maximum amplitude algorithm (MAP) was applied on the oscillation waveforms of the APV model to obtain the mean arterial pressure and the characteristic ratio. The estimation of mean arterial pressure and characteristic ratio was significantly affected with the shape of the blood pressure waveforms and the cutoff frequency of high-pass filter (HPL) circuitry. Experimental errors are due to these effects when estimating blood pressure. To find out an algorithm independent from the influence of waveform shapes and parameters of HPL, the volume oscillation of the APV model and the phase shift of the oscillation with fast Fourier transform (FFT) were testified while increasing the cuff pressure from 1 mmHg to 200 mmHg (1 mmHg per second). The phase shift between the ranges of volume oscillation was then only observed between the systolic and the diastolic blood pressures. The same results were also obtained from the simulations performed on two different the arterial blood pressure waveforms and one hyperthermia waveform.
更多
查看译文
关键词
arterial pressure-volume model,blood pressure,fast fourier transform,manometers,arterial pressure,blood pressure measurement,arterial blood pressure,diastolic blood pressure,high pass filter,high pass filters,computer simulation,phase shift,oscillations,fast fourier transforms
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要