Design of a Compton camera for 3D prompt-γ imaging during ion beam therapy

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2011)

引用 75|浏览16
暂无评分
摘要
We investigate, by means of Geant4 simulations, a real-time method to control the position of the Bragg peak during ion therapy, based on a Compton camera in combination with a beam tagging device (hodoscope) in order to detect the prompt gamma emitted during nuclear fragmentation. The proposed set-up consists of a stack of 2mm thick silicon strip detectors and a LYSO absorber detector. The γ emission points are reconstructed analytically by intersecting the ion trajectories given by the beam hodoscope and the Compton cones given by the camera. The camera response to a polychromatic point source in air is analyzed with regard to both spatial resolution and detection efficiency. Various geometrical configurations of the camera have been tested. In the proposed configuration, for a typical polychromatic photon point source, the spatial resolution of the camera is about 8.3mm FWHM and the detection efficiency 2.5×10−4 (reconstructable photons/emitted photons in 4π). Finally, the clinical applicability of our system is considered and possible starting points for further developments of a prototype are discussed.
更多
查看译文
关键词
Compton camera,Hadrontherapy,Ion beam therapy,Prompt gamma,Geant4
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要