Chrome Extension
WeChat Mini Program
Use on ChatGLM

New Thermoluminescent Dosimeters (tld): Optimization and Characterization of Tld Threads Sterilizable by Autoclave

Physics in medicine & biology/Physics in medicine and biology(2004)

Cited 12|Views8
No score
Abstract
To improve the performance of mono-extruded TLD threads as a dosimetric thermoluminescent tool (French Patent 9903729), a new process was developed by co-extrusion methodology leading to threads of 600 microm diameter with a 50 microm homogeneous polypropylene sheath. In this optimization work, study of parameters such as LiF:Mg,Cu,P powder granulometry, load rate and proportion of components led to an increased sensitivity of around 40%. Moreover, the co-extrusion technique allowed the threads to be sterilized by humid steam (134 degrees C/18 min) without significant variation of the linearity response between 0 and 30 Gy after gamma irradiation (60Co).