Reactivity Of A Dual Amplified Chlamydia Immunoassay With Different Serovars Of Chlamydia Trachomatis

INTERNATIONAL JOURNAL OF STD & AIDS(1999)

引用 10|浏览2
暂无评分
摘要
A study was undertaken with different serovars (D, E, F, L2, MoPn) of Chlamydia trachomatis to determine the analytical sensitivity of a new dual amplified immunoassay (IDELA PCE Chlamydia) for detecting chlamydial lipopolysaccharide. IDEIA PCE Chlamydia incorporates a polymer conjugate consisting of multiple copies of antibody and enzyme molecules to provide signal amplification. The test was also assessed with different protein A producing strains of Staphylococcus aureus in order to assess whether the use of a multiple antibody conjugate increased nonspecific binding. The detection limits varied for each serovar with a detection limit of 38 IFU/ml obtained with serovar F and 237 IFU/ml obtained with serovar D. The incorporation of the polymer conjugate resulted in a 2-5 fold increase in analytical sensitivity compared to an earlier version of the test using a conventional conjugate. No increase in cross reactivity with protein A producing strains of S. aureus was obtained. The new dual amplified test format offers potential as a sensitive low-cost screening assay for C. trachomatis infections.