Data-driven compensation for flow cytometry of solid tissues.

Adv. Bioinformatics(2011)

引用 3|浏览8
暂无评分
摘要
Propidium Iodide is a fluorochrome that is used to measure the DNA content of individual cells, taken from solid tissues, with a flow cytometer. Compensation for spectral cross-over of this fluorochrome still leads to compensation results that are depending on operator experience. We present a data-driven compensation (DDC) algorithm that is designed to automatically compensate combined DNA phenotype flow cytometry acquisitions. The generated compensation values of the DDC algorithm are validated by comparison with manually determined compensation values. The results show that (1) compensation of two-color flow cytometry leads to comparable results using either manual compensation or the DDC method; (2) DDC can calculate sample-specific compensation trace lines; (3) the effects of two different approaches to calculate compensation values can be visualized within one sample. We conclude that the DDC algorithm contributes to the standardization of compensation for spectral cross-over in flow cytometry of solid tissues.
更多
查看译文
关键词
bioinformatics,biomedical research
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper