Comparison of two human immunodeficiency virus (HIV) RNA surrogate assays to the standard HIV RNA assay.

JOURNAL OF CLINICAL MICROBIOLOGY(2005)

引用 48|浏览14
暂无评分
摘要
Human immunodeficiency virus (HIV) RNA testing is the gold standard for monitoring antiretroviral therapy in HIV-infected patients. However, equipment and reagent costs preclude widespread use of the assay in resource-limited settings. The Perkin-Elmer Ultrasensitive p24 assay and the Cavidi Exavir Load assay both offer potentially simpler, less costly technologies for monitoring viral load. These assays were compared to the Roche Amplicor HIV-1 Monitor Test, v1.5, using panels of clinical samples (subtype B) from HIV-positive subjects and HIV-spiked samples (subtypes A, C, D, CRF_01AE, CRF_02AG, and F). The Ultrasensitive p24 assay detected 100% of the spiked samples with virus loads of > 250,000copies/ml and 61% of the clinical samples with virus loads of 219 to 288,850 copies/ml. Detection rates were improved substantially if an external lysis buffer was added to the procedure. The Cavidi assay detected 54 to 100% of spiked samples with virus loads > 10,000 copies/ml and 68% of the clinical samples. These detection rates were also greatly improved with a newly implemented version of this kit. Coefficients of variation demonstrate good reproducibility for each of these kits. The results from the Cavidi v1.0, Cavidi v2.0, and Perkin-Elmer, and the Perkin-Elmer Plus external buffers all correlated well with the results from the Roche Monitor Test (r = 0.83 to 0.96, r = 0.84 to 0.99, r = 0.58 to 0.67, and r = 0.59 to 0.95, respectively). Thus, the use of these two assays for monitoring patients, together with less-frequent confirmation testing, offers a feasible alternative to frequent HIV RNA testing in resource-limited settings.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要