SU-FF-J-84: Shielding of Radio Frequency Noise From Clinical Linear Accelerators for the Linac-MR Project

MEDICAL PHYSICS(2009)

引用 0|浏览1
暂无评分
摘要
Purpose: To determine the efficacy of shielding RFnoise generated by a linac during irradiation. Method and Materials: The electric (E) and magnetic (H) field strengths generated by a medicallinac during irradiation were measured using near field E and H probes connected with a 2 GHz digital oscilloscope and IEEE 488 data acquisition system. These measurements were performed both with and without our RF shield, which houses our 0.2 T MRI. These field measurements were then used to determine the power spectral density of RFnoise as a function of frequency. Results: The data measured without the RF shield shows several peaks in the power spectrum approaching levels of 1mW/m2 in the frequency range 1–20 MHz. Above 5 MHz the measured data with the RF shield shows significantly reduced power levels. For example at 35 MHz the attenuation is 38.5 dB. The RFnoise generated by a linac during the MR imaging k‐space acquisition must be less than μWs. In a 50 kHz bin, our data indicates power densities of ∼250, 69, 7.2 and 2.6 nW/m2 at 8.5, 21.25, 42.5 and 63.75 MHz, corresponding to 0.2, 0.5 1.0 and 1.5 T MR resonant frequencies, respectively. For a typical surface coil used for human imaging, the surface area might be 20×20 cm2. Thus the power available to a typical surface coil, according to our measurements, would be a maximum of 10, 2.76, 0.29 and 0.1 nW respectively. This does not include the coil efficiency which would further reduce the actual RFnoise reaching the acquisition system. Conclusion: The noise power levels measured using our RF shielding is much lower than those detected in typical human MR imaging.
更多
查看译文
关键词
radio frequency,linear accelerator
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要