Purpose: To investigate diagnostic accuracy of detection of prostate cancer by magnetic resonance: to evaluate the performance of T2WI, DCEMRI and CSI and to correlate the results with biopsy and radical prostatectomy histopathological data.Materials and methods: 43 patients, scheduled for radical prostatectomy, underwent prostate MR examination. Prostate cancer was identified by transrectal ultrasonographically (TRUS) guided sextant biopsy. MR examination was performed at 1.5T with an endorectal MR coil. Cancer localisation was performed on sextant-basis - for comparison between TRUS biopsy, MR techniques and histopathological findings on prostatectomy specimens.Results: Prostate cancer was identified in all 43 patients by combination of the three MR techniques. The detection of prostate cancer on sextant-basis showed sensitivity and specificity: 50% and 91% for TRUS, 72% and 55% for T2WI, 49% and 69% for DCEMRI, and 46% and 78% for CSI.Conclusion: T2WI, DCEMRI and CSI in combination can identify prostate cancer. Further development of MR technologies for these MR methods is necessary to improve the detection of the prostate cancer. (C) 2010 Elsevier Ireland Ltd. All rights reserved.