Aldosterone quantification helps evaluate the rennin-angiotensin-aldosterone system. The new bead-based mul-tiplex platform has not been applied in aldosterone detection to achieve simultaneous measurements of multiple hormones. A new sensitive competitive bead immunoassay based on Luminex technology for detecting aldoster-one in small sample volumes was developed using two-antibody coupled beads and biotinylated aldosterone as tracer in combination with an extraction step. The assay was validated in human and mouse samples and exhibited a linear working range from 10 to 1,000 pg/mL. The assay was reproducible and precise with intra-assay coefficient of variations (CVs) from 6.0% to 11.2%, inter-assay CVs from 8.0% to 13.0% and good recovery [(90-110)%] and linearity [(89-107)%]. Excellent correlation was found between this new assay and the reference method (r = 0.96, P < 0.000,1). The successful establishment of this assay provides high possibility for carrying out bead-based multiplex assay measuring aldosterone and other parameters simultaneously in one 50 μL sample so that the efficiency can be improved and precious samples can be saved.
More
Translated text
Key words
aldosterone,immunoassay,bead,extraction,multiplex