Direkte Autotransfusionssysteme liefern Blut unzureichender Qualität

ANASTHESIOLOGIE INTENSIVMEDIZIN NOTFALLMEDIZIN SCHMERZTHERAPIE(2000)

引用 20|浏览3
暂无评分
摘要
Systems for direct retransfusion of blood salvaged from the surgical field and in drainage systems (direct autotransfusion) have been in use for many years. The quality of the blood obtained with such systems, however, has not been systemically assessed in a comparative manner. The aim of our study was the analysis of the quality of the blood, obtained with three commercially available direct autotransfusion systems (drainage systems with filters). Methods: With ethics committee approval and informed consent, 30 patients receiving knee arthroplasty were randomly allocated to three groups. Each group of 10 patients received treatment with one drainage system (Consta Vac(TM), Solcotrans(TM), Haem-o-Trans(TM)). In the salvaged blood, we measured cellular elements, variables of coagulation and fibrinolysis, complement activation and cellular elements, both before and after passage of the autotransfusion system. Results: Analysis revealed a low haematokrit (< 30 %) and platelet count (< 80 Gpt/l). The salvaged blood proved uncoagulable and defibrinised with no measurable clotting and fibrinogen; dotting activity, fibrinolysis and complement reaction were grossly induced (TAT, PAP and FDP high, C-3 low). The blood was contaminated with cellular debris reflected by concentration of enzymes usually confined to the intracellular space (LDH, elastase, beta-thromboglobulin). Conclusion: The systems/filters assessed in this study do not improve quality of blood drained from the surgical field. Retransfusion of such blood can not be recommended.
更多
查看译文
关键词
transfusion,autotransfusion systems,quality blood,drainage systems,orthopaedic patients
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要