Charge transport and signal generation in CdTe pixel detectors

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(1996)

引用 39|浏览12
暂无评分
摘要
The current and charge signals on individual pixels of CdTe γ-ray cameras are calculated. The time and position dependent free carrier densities following the absorption of a γ-ray in the detector are first calculated by solving the continuity equations. These density distributions are then combined with the physical electric field and the pixel weighting field to provide the current induced through an individual pixel from which the charge signal is also obtained. Simulations are made for various pixel sizes. These simulations confirm that the energy resolution of electron collecting pixel detectors must improve when smaller pixels are used since small pixels are less sensitive to holes.
更多
查看译文
关键词
continuity equation,electric field
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要