谷歌浏览器插件
订阅小程序
在清言上使用

Simple Real Time Trace Nitrogen Dioxide Detector Based on Continuous-Wave Cavity Ringdown Spectroscopy Using Passively Locked External Cavity Diode Laser

JAPANESE JOURNAL OF APPLIED PHYSICS(2008)

引用 7|浏览1
暂无评分
摘要
We report on a compact trace nitrogen dioxide (NO) detector based on cavity ringdown spectroscopy (CRDS), using a passively locked external cavity diode laser (PLEC-DL) in the visible wavelength region. Tailoring the spatial mode matching conditions of an antireflection (AR)-coated diode laser with a high finesse external cavity for effective optical feedback into the DL allows for stable resonant enhancement in the external cavity without any electric feedback apparatus. The external cavity having a finesse in excess of 1.4 x 10(5) acts as a ringdown cavity with an effective optical volume (i.e., detection volume) of less than 1 cm(3). We successfully demonstrated a noise equivalent absorption loss of 1.4 x 10(-8)cm(-1), corresponding to a detection limit of 40 ppb NO(2) concentration with a very compact configuration. For differential absorption CRDS, we realized a dual wavelength-locked PLEC-DL in an external cavity simultaneously. Measurements of various NO(2) concentrations in N(2) based on differential absorption CRDS were performed.
更多
查看译文
关键词
laser spectroscopy,external cavity,diode laser,cavity ringdown spectroscopy,nitrogen dioxide detection
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要