Performance of a medical imaging system for photons in the 60–140 keV energy range

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2001)

引用 9|浏览10
暂无评分
摘要
We report the status of the art of a prototype based on a GaAs pixel detector bump-bonded to a dedicated VLSI chip to be possibly used for imaging in the nuclear medicine field. This device, with a 200 mum thick pixel matrix (64 x 63 square pixels, 170 mum side), has already been tested with very good results for digital mammography applications (mean energy 20 keV). For more energetic photons, as in nuclear medicine, a 600 mum thick detector has been chosen. Using radioactive sources (Am-241, 60 keV and Tc-99m, 140 keV photons) we have measured the performance of our pr otatype in terms of charge collection and detection efficiency of the detector, discrimination capability of the electronics and imaging properties of the whole system. In particular, we have evaluated the spatial resolution properties measuring the Point Spread Function and the imaging capabilities using a home made thyroid phantom. We present also the comparison between these results and those obtained with a traditional gamma camera and the evaluation, made by both experimental measurements and software simulations, of the geometry related to the use of a collimator. (C) 2001 Elsevier Science B.V. All rights reserved.
更多
查看译文
关键词
nucler medicine,GaAs detector,single photon counting