Cold Performance Tests Off Blocked-Impurity-Band Si:As Detectors Developed For Darwin

HIGH ENERGY, OPTICAL, AND INFRARED DETECTORS FOR ASTRONOMY III(2008)

引用 3|浏览22
暂无评分
摘要
We report first results of laboratory tests of Si:As blocked-impurity-band (BIB) mid-infrared (4 to 28 mu m) detectors developed by IMEC. These prototypes feature 88 pixels hybridized on an integrated cryogenic readout electronics (CRE). They were developed as part of a technology demonstration program for the future DARWIN mission. In order to be able to separate detector and readout effects, a custom build TIA circuitry was used to characterize additional single pixel detectors. We used a newly designed test setup at the MPIA to determine the relative spectral response, the quantum efficiency, and the dark current. All these properties were measured as a function of operating temperature and detector bias. In addition the effects of ionizing radiation on the detector were studied. For determining the relative spectral response we used a dual-grating monochromator and a bolometer with known response that was operated in parallel to the Si:As detectors. The quantum efficiency was measured by using a custom-build high-precision vacuum black body together with cold (T similar to 4 K) filters of known (measured) transmission.
更多
查看译文
关键词
mid-infrared detectors, Si:As blocked-impurity-band detectors, detector characterization, dark current, responsivity, relative spectral response, DARWIN