Increased incidence of gastrointestinal bleeding following implantation of the HeartMate II LVAD.

The Journal of Heart and Lung Transplantation(2010)

引用 198|浏览3
暂无评分
摘要
P>Background: The HeartMate II (HMII) Left Ventricular Assist System (Thoratec Corporation, Pleasanton, CA, USA), an axial continuous-flow left ventricular assist device (LVAD), has been approved for use in bridge-to-transplant patients and is under investigation for destination therapy. To avoid device-related thromboembolic complications, antiplatelet, and anticoagulation therapy are routinely administered. A worrisome frequency of gastrointestinal (GI) bleeding events has been observed. Methods: A retrospective review of all 33 patients undergoing long-term LVAD implantation between June 1, 2006 and July 31, 2008 at our institution for any indication was conducted. Anticoagulation consisted of heparin (intravenous or subcutaneous) followed by transition to Coumadin therapy to a target INR of two to three. Antiplatelet therapy consisted of low-dose aspirin and dipyridamole. Results: Twenty patients received the HMII and 13 patients received other devices. Eight (40%) HMII recipients suffered at least one episode of GI bleeding while no GI bleeding occurred in recipients of other devices (p = 0.012). Of 17 total bleeding episodes, no definitive source could be identified in 11 instances (65%). Conclusions: Although definitive source identification remains elusive, we believe that the majority of bleeding arises in the small bowel, possibly due to angiodysplasias, similar to the pathophysiology encountered in patients with aortic stenosis and GI bleeding. As we move toward wider use of the HMII and other axial continuous-flow devices in both bridge-to-transplant patients and for destination therapy, more studies will be necessary to understand the mechanisms of this obscure GI bleeding and develop treatment strategies to minimize its development. (J Card Surg 2010;25:352-356).
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要