谷歌浏览器插件
订阅小程序
在清言上使用

Effects of Fluid Viscoelasticity on the Performance of an Axial Blood Pump Model

ASAIO journal(2012)

引用 1|浏览12
暂无评分
摘要
An aqueous Xanthan gum solution (XGS) was used as blood analog fluid to explore the influence of fluid viscoelasticity on the performance of an axial blood pump model. For comparison, a 39 wt% Newtonian aqueous glycerin solution (GS), the common fluid in blood pump tests, was also used as a working fluid. The experimental results showed that a higher head curve was obtained using XGS in the pump than using GS. The heads of the XGS that were computed using the viscoelastic turbulence model agreed well with the measured data. In contrast, the standard k-&egr; turbulence model failed to provide satisfactory predictions for the XGS. The computational results revealed that in most parts of the pump model flow fields, the Reynolds shear stress values and turbulent dissipation rates of the XGS were all lower than those of the GS. The hemolysis index of the pump model using the XGS was calculated to be only one-third of that using the GS.