Dosimetric experience with 2 commercially available multilumen balloon-based brachytherapy to deliver accelerated partial-breast irradiation.

Medical Dosimetry(2015)

引用 2|浏览8
暂无评分
摘要
The purpose of this work was to report dosimetric experience with 2 kinds of multilumen balloon (MLB), 5-lumen Contura MLB (C-MLB) and 4-lumen MammoSite MLB (MS-MLB), to deliver accelerated partial-breast irradiation, and compare the ability to achieve target coverage and control skin and rib doses between 2 groups of patients treated with C-MLB and MS-MLB brachytherapy. C-MLB has 5 lumens, the 4 equal-spaced peripheral lumens are 5mm away from the central lumen. MS-MLB has 4 lumens, the 3 equal-spaced peripheral lumens are 3mm away from the central lumen. In total, 43 patients were treated, 23 with C-MLB, and 20 with MS-MLB. For C-MLB group, 8 patients were treated with a skin spacing < 7mm and 12 patients with rib spacing < 7mm. For MS-MLB group, 2 patients were treated with a skin spacing < 7mm and 5 patients with rib spacing < 7mm. The dosimetric goals were (1) ≥ 95% of the prescription dose (PD) covering ≥ 95% of the target volume (V95% ≥ 95%), (2) maximum skin dose ≤ 125% of the PD, (3) maximum rib dose ≤ 145% of the PD (if possible), and (4) the V150% ≤ 50cm3 and V200% ≤ 10cm3. All dosimetric criteria were met concurrently in 82.6% of C-MLB patients, in 80.0% of MS-MLB patients, and in 81.4% of all 43 patients. For each dosimetric parameter, t-test of these 2 groups showed p > 0.05. Although the geometric design of C-MLB is different from that of MS-MLB, both applicators have the ability to shape the dose distribution and to provide good target coverage, while limiting the dose to skin and rib. No significant difference was observed between the 2 patient groups in terms of target dose coverage and dose to organs at risk.
更多
查看译文
关键词
Balloon brachytherapy,Accelerated partial-breast irradiation,Breast cancer,Contura MLB,MammoSite MLB
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要