Whole brain, high resolution multiband spin-echo EPI fMRI at 7 T: A comparison with gradient-echo EPI using a color-word Stroop task

NeuroImage(2014)

引用 37|浏览20
暂无评分
摘要
A whole brain, multiband spin-echo (SE) echo planar imaging (EPI) sequence employing a high spatial (1.5mm isotropic) and temporal (TR of 2s) resolution was implemented at 7T. Its overall performance (tSNR, sensitivity and CNR) was assessed and compared to a geometrically matched gradient-echo (GE) EPI multiband sequence (TR of 1.4s) using a color-word Stroop task. PINS RF pulses were used for refocusing to reduce RF amplitude requirements and SAR, summed and phase-optimized standard pulses were used for excitation enabling a transverse or oblique slice orientation. The distortions were minimized with the use of parallel imaging in the phase encoding direction and a post-acquisition distortion correction. In general, GE-EPI shows higher efficiency and higher CNR in most brain areas except in some parts of the visual cortex and superior frontal pole at both the group and individual-subject levels. Gradient-echo EPI was able to detect robust activation near the air/tissue interfaces such as the orbito-frontal and subcortical regions due to reduced intra-voxel dephasing because of the thin slices used and high in-plane resolution.
更多
查看译文
关键词
BOLD,PINS,Orbito-frontal,Sensitivity,CNR,Physiological noise
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要