CMOS potentiostat and sensor with multilayer membrane for wide range measurements of glucose concentrations

Valencia(2014)

引用 1|浏览2
暂无评分
摘要
We report on an electrochemical measurement setup comprising a CMOS potentiostat and a glucose sensor with a two-layer membrane as the first steps towards the development of an integrated in-situ sensor system for bioreactors. The potentiostat has a chip size of 2.1 × 2.5 mm2 and a linear current range from -220 nA to 240 nA with a linearity of R2 = 0.9995. For wide range measurements of glucose concentrations in cell culture media, electrodes functionalized with the enzyme glucose oxidase were spin-coated with membranes made from polydimethylsiloxane (PDMS). With these membranes, glucose concentrations up to 400 mM were measured with a linear measurement range up to 100 mM. For interferent elimination, cellulose acetate membranes were employed.
更多
查看译文
关键词
CMOS integrated circuits,bioreactors,chemical sensors,enzymes,membranes,spin coating,sugar,CMOS potentiostat,bioreactors,cellulose acetate membranes,current -220 nA to 240 nA,electrodes,enzyme glucose oxidase,glucose concentrations,glucose sensor,integrated in-situ sensor system,multilayer membrane,polydimethylsiloxane,spin-coating,two-layer membrane,CMOS potentiostat,PDMS,bioreactor monitoring,cellulose acetate,glucose sensor,high glucose concentrations,membrane,wide range measurement