Spacing analysis of a focal magnetic stimulating coil

Durham, NH(1997)

引用 2|浏览2
暂无评分
摘要
A promising coil geometry for focal stimulation of neurons through magnetic induction is the quad-square coil. This configuration utilizes four 90° corners at the center to produce large values of dEx /dx (considered the activating function in neural stimulation). The values of dEx/dx produced by the center of the coil are limited by the coil dimensions and coil spacing. A closed form solution of the limiting function in terms of coil dimension, spacing, and stimulating depth is presented and the relationship between dEx /dx and spacing is graphically shown
更多
查看译文
关键词
biological effects of fields,biological techniques,biomagnetism,biomedical equipment,coils,electromagnetic induction,focusing,neurophysiology,activating function,closed form solution,coil dimensions,coil geometry,coil spacing,focal magnetic stimulating coil,focal stimulation,limiting function,magnetic induction,neural stimulation,neurons,quad-square coil,spacing analysis,stimulating depth,activation function,geometry,permeability,biomedical engineering