The Characterisation of a Secondary Fluorescence X-ray Tube for Medical Imaging, Security Screening and Analytical Applications

Dermody, G.S.,Carter, T., Jupp, I.D., Hunter, S.J.

Nuclear Science Symposium Conference Record, 2000 IEEE(2000)

引用 0|浏览5
暂无评分
摘要
Describes the characterisation of a new prototype secondary fluorescence X-ray tube. This tube allows the user to interchange the secondary target, and thus select the X-ray tube spectrum. Measurements of the focal spot size, intensity the K/spl alpha/ fluorescence lines and the spectral purity of the source are presented for a range of secondary targets including tantalum, tungsten, silver, copper and a rhodium layer on tungsten. Measurements with the tantalum target confirmed that the intensity from the new Fluorex tube was of the same order as measured previously (8.5/spl times/10/sup 5/ photons/s/mm/sup 2/ source area into 1/spl times/10/sup -4/ sr) at 160 kV and 3.75 kW. This was found to be approximately 30-40 times less than the flux into a comparable energy window containing the K/spl alpha/ emission lines of a conventional tungsten anode X-ray tube operating at 140 kV and 3.75 kW.
更多
查看译文
关键词
X-ray tubes,biomedical equipment,diagnostic radiography,fluorescence,140 kV,3.75 kW,Ag,Cu,Fluorex tube,Kα emission lines,Rh,Ta,X-ray tube spectrum selection,analytical applications,conventional tungsten anode X-ray tube,energy window,focal spot size,medical diagnostic imaging,medical imaging,medical instrumentation,rhodium layer,secondary fluorescence X-ray tube characterisation,secondary targets,security screening,source spectral purity