A 10-Fr ultrasound catheter with integrated micromotor for4-D intracardiac echocardiography

Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions(2011)

引用 21|浏览3
暂无评分
摘要
We developed prototype real-time 3-D intracardiac echocardiography catheters with integrated micromotors, allowing internal oscillation of a low-profile 64-element, 6.2-MHz phased-array transducer in the elevation direction. Components were designed to facilitate rotation of the array, including a low-torque flexible transducer interconnect and miniature fixtures for the transducer and micromotor. The catheter tip prototypes were integrated with two-way deflectable 10-Fr catheters and used in in vivo animal testing at multiple facilities. The 4-D ICE catheters were capable of imaging a 90° azimuth by up to 180° elevation field of view. Volume rates ranged from 1 vol/sec (180° elevation) to approximately 10 vol/sec (60° elevation). We successfully imaged electrophysiology catheters, atrial septal puncture procedures, and detailed cardiac anatomy. The elevation oscillation enabled 3-D visualization of devices and anatomy, providing new clinical information and perspective not possible with current 2-D imaging catheters.
更多
查看译文
关键词
bioelectric phenomena,biomedical transducers,catheters,echocardiography,micromotors,physiological models,ultrasonic transducer arrays,3D visualization,4D intracardiac echocardiography,atrial septal puncture procedure,cardiac anatomy,electrophysiology catheters,frequency 6.2 MHz,in-vivo animal testing,integrated micromotor,low-torque flexible transducer interconnect,miniature fixtures,phased-array transducer,prototype real-time 3D intracardiac echocardiography catheters,two-way deflectable catheters,ultrasound catheter
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要