Sedimentation velocity analytical ultracentrifugation and SEDFIT/c(s): limits of quantitation for a monoclonal antibody system.

Analytical Biochemistry(2007)

引用 57|浏览1
暂无评分
摘要
Sedimentation velocity analytical ultracentrifugation (SV–AUC) has emerged in the biopharmaceutical industry as a technique to detect small quantities of protein aggregates. However, the limits of detection and quantitation of these aggregates are not yet well understood. Although diverse factors (molecule, instrument, technique, and software dependent) preclude an all-encompassing measurement of these limits for the complete system, it is possible to use simulated data to determine the quantitation limits of the data analysis software aspect. The current study examines the performance of the SEDFIT/c(s) data analysis tool with simulated antibody monomer/dimer and monomer/aggregate systems. Under completely ideal conditions (zero noise, known meniscus, and shape factor homogeneity), the software limit of quantitation was 0.01% for the monomer/aggregate system and 0.03% for the less well-resolved monomer/dimer system. Under more realistic conditions (0.005 OD root mean square [RMS] noise, shape factor variability, and long solution column), the software limits of quantitation were 0.2 and 0.6% (0.002 and 0.006 OD) for the monomer/aggregate and monomer/dimer systems, respectively. Interestingly, diminished quantitation accuracy at very low levels of oligomer was not accompanied by deterioration of fit quality (as measured by root mean square deviation [RMSD] and residuals bitmap images).
更多
查看译文
关键词
Analytical ultracentrifugation,Sedimentation velocity,SEDFIT,Monoclonal antibody,Limit of quantitation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要