Conductance artifacts in a novel in vitro model of ventriculothoracic electrical coupling.

ASAIO JOURNAL(1999)

引用 5|浏览2
暂无评分
摘要
The utility of open chest conductance (COND) ventriculography is limited by artifacts altering the relationship between COND and left ventricular (LV) volume. Pressure-COND loops often lean to the left during LV volume reduction by caval occlusion. Time varying alterations in the pericardial-LV contact area affect electrical coupling in the open chest during the cardiac cycle, producing COND artifacts. In this study, an open-mediastinum model was constructed. Components represented the LV, blood, pericardium, and thoracic contents. Varying ventriculothoracic coupling was simulated by changing the volume of pericardial saline (0, 30, 60 ml). Raw dual field COND was repeatedly (n = 20) compared with volumes of normal saline from 60 to 120 ml at 5 ml intervals. Groups were compared by linear regression and repeated measures ANOVA. Artifacts significantly (p < 0.01) altered parallel COND, indicated by the y-intercept, with the exception of 0 verses 30 ml. The slope constant also changed significantly, with the exception of 30 versus 60 ml. These results suggest that variable pericardial-LV contact can cause time varying artifacts in COND in the open chest. Therefore, posterior insulation may reduce artifacts in COND ventriculography and should be tested for this effect.