Comparison of dynamic contour tonometry and Goldmann applanation tonometry in keratoconus.

CORNEA(2011)

引用 13|浏览3
暂无评分
摘要
Purpose: We performed a comparative study using Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) to measure intraocular pressure (IOP) in eyes with keratoconus. Methods: IOP was measured in 114 eyes of 75 patients with keratoconus (51 men and 24 women; mean age, 36.1 +/- 11.2 years) using GAT and DCT in randomized order. Central corneal thickness (CCT), minimal corneal thickness (MCT), and corneal topography were recorded using the Pentacam. Four groups according to Amsler's keratoconus classification were composed and analyzed for significant differences of CCT, MCT, GAT, and DCT results. Results: Mean CCT in the 114 keratoconus eyes was 481.1 +/- 46.2 mu m (range, 334-601 mu m). Mean MCT was 453.3 +/- 56.3 mu m (range, 239-573 mu m). Mean IOP measured using GAT was 13.1 +/- 2.9 mm Hg, whereas mean IOP measured using DCT was 14.8 +/- 2.6 mm Hg. Neither the results for GAT nor those for DCT showed a significant correlation with CCT (Pearson correlation: P < 0.05). Multifactorial analysis revealed that CCT and MCT, but not GAT and DCT, results were significantly different in corneas of varying curvatures. Conclusions: This study shows that DCT measures IOP higher than GAT in eyes with keratoconus. In keratoconus, both methods seem to be independent of CCT and therefore are equally, but not interchangeably, applicable when monitoring IOP. Further analysis revealed that CCT and MCT are significantly different in corneas of varying Amsler grade.
更多
查看译文
关键词
keratoconus,intraocular pressure,Goldmann applanation tonometry,dynamic contour tonometry,glaucoma
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要