Improvement of multislice oxygen-enhanced MRI of the lung by fully automatic non-rigid image registration.
European Journal of Radiology(2012)
摘要
In oxygen-enhanced magnetic resonance imaging of the lung (O2-MRI), motion artifacts related to breathing hamper the quality of the parametric O2-maps. In this study, fully automatic non-rigid image registration was assessed as a post-processing method to improve the quality of O2-MRI.Twenty healthy volunteers were investigated on a 1.5 T MR system. O2-MRI was obtained in four coronal sections using an IR-HASTE sequence with TE/TI of 12/1200 ms. Each section was repeatedly imaged during oxygen and room-air ventilation. Spatial differences among the images were corrected by fully automatic non-rigid registration. Signal variability, relative enhancement ratio between oxygen and room air images, and spatial heterogeneity of lung enhancement were assessed before and after image registration.Motion artifacts were corrected in 5-10s. Non-rigid registration reduced signal variability of the source images and heterogeneity of the O2-maps by 1.1 ± 0.2% and 11.2 ± 2.9%, respectively (p<0.0001). Registration did not influence O2 relative enhancement ratio (p=0.06).Fully automatic non-rigid image registration improves the quality of multislice oxygen-enhanced MRI of the lung.
更多查看译文
关键词
Lung,Magnetic resonance imaging,Oxygen,Image processing
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要